E-cylinders are used for delivery of fresh gases to the anesthesia machine in case of pipeline gas pressure failure. | E-cylinders are used for delivery of fresh gases to the anesthesia machine in case of pipeline gas pressure failure. | ||
E-cylinders | E-cylinders uses the pin index safety system (PISS) that ensures the correct placement of cylinders of a particular type of gas into the right location in the anesthesia machine. | ||
{| class="wikitable" | {| class="wikitable" | ||
|+ | |+ |
Revision as of 07:33, 9 August 2021
E-cylinders are used for delivery of fresh gases to the anesthesia machine in case of pipeline gas pressure failure.
E-cylinders uses the pin index safety system (PISS) that ensures the correct placement of cylinders of a particular type of gas into the right location in the anesthesia machine.
Gas | Cylinder Color (USA) | Full Tank Volume (L) | Full Tank Pressure (PSI) |
---|---|---|---|
Oxygen | Green | 660 | 2200 |
Nitrous Oxide | Blue | 750 | 1590 |
Air | Yellow | 625 | 1800 |